Development of Chemiresistor Based Nanosensors to Detect Volatile Cancer Biomarkers

Download (10.84 MB)
thesis
posted on 2021-10-12, 14:44 authored by Shitiz VijShitiz Vij
Researchers have shown links between various hydrocarbons and carbonyl compounds and diseases, such as cancer using exhaled breath analysis through gas chro
matography/mass spectroscopy (GC/MS) analysis of volatile organic compounds (VOCs). Trained canines can detect these VOCs and can differentiate a patient suffering from cancer from a healthy control patient. In this project, an attempt has been made to develop highly sensitive sensors for the detection of low concentrations of aldehyde VOCs, such as nonanal, using conductive polymer composites (CPCs) and functionalized gold nanoparticles (f-GNPs). Facile methods have been used to enhance the sensitivity and cross-selectivity of the fabricated sensors towards nonanal. Interdigitated electrodes (IDEs) are fabricated through a photolithography process. Sensors of PEI/carbon black (CB) composite were developed via spin-coating of the
material followed by the heat treatment process. Sensors of 1-Mercapto-(triethyleneglycol) methyl ether functionalized GNPs are developed via drop-casting of nanomaterial and f-GNP/PEI sensors are fabricated by spin casting PEI film on top of f-GNPs. Fourier Transform Infrared (FTIR) analysis, X-Ray Diffraction (XRD) analysis, contact angle measurement, and Field Emission Scanning Electron Microscopy (FESEM) analysis was conducted to characterize the fabricated devices. The fabricated sensors have been tested with a low concentration of nonanal, nonanone, dodecane, and 1-octanol in dry air. Multiple sensors are fabricated to ensure sensors reproducibility. The sensors have been exposed repeatedly to the targeting VOCs to assess the repeatability of the sensors. PEI/CB sensor degradation was studied over a period of 36 days.

The fabricated PEI/CB film could detect (1-80 ppm) of nonanal with higher selectivity, than the f-GNPs. The sensor0s sensitivity to nonanal was over fourteen times
higher than 2-nonanone, 1-octanol, and dodecane. This shows the high selectivity of the fabricated sensor toward nonanal. In addition, the proposed sensor maintained its
sensitivity to nonanal over time showing minimal degradation. The sensor response to nonanal at a relative humidity (RH) of 50% and 85% dropped less than 13% and
32% respectively. The Response of f-GNP sensors to nonanal (400 ppb - 15 ppm), dodecane (5 - 15 ppm), 1-octanol (5 - 15 ppm), and 2-nonanone (5 - 15 ppm) presented a sensitivity (∆R/R0) of 0.217%, 0.08%, 0.192% and 0.182% per ppm of the VOCs respectively. Despite the high sensitivity to the targeting VOCs, the fabricated
sensors were damaged in an environment with relative humidity (RH) at 45%. A thin layer of PEI over the film was developed to ensure the sensor could tolerate long
time exposure to water vapor in an environment with RH up to 85% and enhance the sensor selectivity towards nonanal. The f-GNP/PEI sensors with nonanal (400 ppb- 15 ppm), dodecane (100 -200 ppm), 1-octanol (5 - 15 ppm) and 2-nonanone (5 - 15 ppm) presented sensitivity (∆R/R0) of 0.21%, 0.017%, 0.0438% and 0.0035% per ppm of the VOCs respectively. The sensor fabrication, characterization, testing methods, and results are presented and discussed.

Funding

SCH: EXP: Canine-Inspired Smart Sensor for Detecting Hypoglycemia from Human Breath

Directorate for Computer & Information Science & Engineering

Find out more...

History

Degree Type

  • Master of Science in Mechanical Engineering

Department

  • Mechanical Engineering

Campus location

  • Indianapolis

Advisor/Supervisor/Committee Chair

Mangilal Agarwal

Additional Committee Member 2

Sohel Anwar

Additional Committee Member 3

Andres Towar

Usage metrics

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC